Procedure for preparation and standardization 0.05 N Ceric Ammonium

1. OBJECTIVE
To lay down the procedure for preparation and standardization of 0.05 N Ceric Ammonium Nitrate

2. SCOPE
This procedure is applicable to Quality Control Department

3. RESPONSIBILITY
• Officer QC
• Manager QC

4. SAFETY INSTRUCTIONS
• Proper handling the apparatus.
• Proper handling the chemicals

5. PROCEDURE
5.1 PREPARATION
5.1.1 Weigh out 2.75 g of ceric ammonium nitrate using a balance.
5.1.2 Dissolve 2.75 g of ceric ammonium nitrate in 1 N nitric acid to obtain 100 mL of solution, and filter.
5.1.3 The resulting solution is a 0.05 N solution of ceric ammonium nitrate.
5.2 STANDARDIZATION
5.2.1 Accurately measure 10 mL of freshly standardized 0.1 N ferrous ammonium sulfate VS into a flask,
5.2.2 And dilute with water to about 100 ml. Add 1 drop of nitrophenanthroline TS, and titrate with the ceric ammonium nitrate solution to a colorless endpoint.
5.2.3 And calculate the normality by the following formula:
ml Fe (NH4)2 x N Fe (NH4)2 (SO4)2
N= —————————————————–
ml of Ce(NO3)4 . 2NH4NO3

6. RELATED DOCUMENTS
N/A

7. DISTRIBUTION INDEX
Original Copy will be retained by Quality Assurance Department.

DISTRIBUTION INDEX
Distributed To Copy No. Received By/Date
Quality Control Department 01 ,

8. ABBREVIATIONS
Q.C Quality Control Department
TS Test Solution
Vs. Against
Gm Gram
ml Milliliter

9. REVISION HISTORY

Revision No. Issue Date Supersedes Reason of Change / Description
          Revision No.              Dated     
00 10/02/2023 00                            N/A First Version